<?xml version="1.0" encoding="UTF-8"?><rss version="2.0"
	xmlns:content="http://purl.org/rss/1.0/modules/content/"
	xmlns:wfw="http://wellformedweb.org/CommentAPI/"
	xmlns:dc="http://purl.org/dc/elements/1.1/"
	xmlns:atom="http://www.w3.org/2005/Atom"
	xmlns:sy="http://purl.org/rss/1.0/modules/syndication/"
	xmlns:slash="http://purl.org/rss/1.0/modules/slash/"
	>

<channel>
	<title>Ventricular Fibrillation Archives - EMFSA</title>
	<atom:link href="https://www.emfsa.co.za/tag/ventricular-fibrillation/feed/" rel="self" type="application/rss+xml" />
	<link>https://www.emfsa.co.za/tag/ventricular-fibrillation/</link>
	<description>Electromagnetic fields South Africa</description>
	<lastBuildDate>Fri, 16 Apr 2021 20:02:37 +0000</lastBuildDate>
	<language>en-US</language>
	<sy:updatePeriod>
	hourly	</sy:updatePeriod>
	<sy:updateFrequency>
	1	</sy:updateFrequency>
	<generator>https://wordpress.org/?v=7.1.3</generator>

<image>
	<url>https://www.emfsa.co.za/wp-content/uploads/2021/02/cropped-EMFSA_logo-fv-32x32.png</url>
	<title>Ventricular Fibrillation Archives - EMFSA</title>
	<link>https://www.emfsa.co.za/tag/ventricular-fibrillation/</link>
	<width>32</width>
	<height>32</height>
</image> 
	<item>
		<title>Pacemaker-induced ventricular fibrillation during radiofrequency catheter ablation for ventricular tachycardia</title>
		<link>https://www.emfsa.co.za/research-and-studies/pacemaker-induced-ventricular-fibrillation-during-radiofrequency-catheter-ablation-for-ventricular-tachycardia/</link>
		
		<dc:creator><![CDATA[Editor]]></dc:creator>
		<pubDate>Fri, 16 Apr 2021 20:02:35 +0000</pubDate>
				<category><![CDATA[Research and Studies]]></category>
		<category><![CDATA[Cardiac]]></category>
		<category><![CDATA[CIED]]></category>
		<category><![CDATA[Electromagnetic Interference]]></category>
		<category><![CDATA[Radiofrequency Catheter Ablation]]></category>
		<category><![CDATA[Ventricular Fibrillation]]></category>
		<guid isPermaLink="false">https://www.emfsa.co.za/?p=20507</guid>

					<description><![CDATA[<p>Nam MCY, Jones N, Claridge S, Balasubramaniam R, Sopher M, Babu G. Pacemaker-induced ventricular fibrillation during radiofrequency catheter ablation for ventricular tachycardia. J Arrhythm. 2021 Feb 14;37(2):368-369. doi: 10.1002/joa3.12517. PMID: 33850578; PMCID: PMC8021979. Abstract Prior to ventricular tachycardia ablation, this patient&#8217;s cardiac implantable electronic device (CIED) was temporarily programmed to backup pacing mode with tachycardia [&#8230;]</p>
<p>The post <a href="https://www.emfsa.co.za/research-and-studies/pacemaker-induced-ventricular-fibrillation-during-radiofrequency-catheter-ablation-for-ventricular-tachycardia/">Pacemaker-induced ventricular fibrillation during radiofrequency catheter ablation for ventricular tachycardia</a> appeared first on <a href="https://www.emfsa.co.za">EMFSA</a>.</p>
]]></description>
										<content:encoded><![CDATA[
<p class="wp-block-paragraph" style="font-size:14px"><strong>Nam MCY, Jones N, Claridge S, Balasubramaniam R, Sopher M, Babu G. Pacemaker-induced ventricular fibrillation during radiofrequency catheter ablation for ventricular tachycardia. J Arrhythm. 2021 Feb 14;37(2):368-369. doi: 10.1002/joa3.12517. PMID: 33850578; PMCID: PMC8021979.</strong></p>



<h2 class="wp-block-heading" style="font-size:14px">Abstract</h2>



<p class="wp-block-paragraph" style="font-size:14px">Prior to ventricular tachycardia ablation, this patient&#8217;s cardiac implantable electronic device (CIED) was temporarily programmed to backup pacing mode with tachycardia therapies disabled. During radiofrequency energy delivery, the patient developed ventricular fibrillation requiring emergent cardioversion. Electrogram interrogation showed that the CIED switched to noise reversion mode during ablation. The consequent asynchronous pacing resulted in a paced QRS landing on an intrinsic T wave, inducing ventricular fibrillation. This serves as an important reminder that asynchronous pacing consequent to CIED oversensing could occur in any procedure that could cause electromagnetic interference such as radiofrequency catheter ablation.</p>



<p class="wp-block-paragraph" id="copyright" style="font-size:14px">© 2021 The Authors. Journal of Arrhythmia published by John Wiley &amp; Sons Australia, Ltd on behalf of the Japanese Heart Rhythm Society.</p>



<p class="wp-block-paragraph" style="font-size:14px"><a href="https://pubmed.ncbi.nlm.nih.gov/33850578/">https://pubmed.ncbi.nlm.nih.gov/33850578/</a></p>
<p>The post <a href="https://www.emfsa.co.za/research-and-studies/pacemaker-induced-ventricular-fibrillation-during-radiofrequency-catheter-ablation-for-ventricular-tachycardia/">Pacemaker-induced ventricular fibrillation during radiofrequency catheter ablation for ventricular tachycardia</a> appeared first on <a href="https://www.emfsa.co.za">EMFSA</a>.</p>
]]></content:encoded>
					
		
		
			</item>
	</channel>
</rss>
